These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Real-time quadrature projection complex conjugate resolved Fourier domain optical coherence tomography. Author: Sarunic MV, Applegate BE, Izatt JA. Journal: Opt Lett; 2006 Aug 15; 31(16):2426-8. PubMed ID: 16880844. Abstract: We present a novel algorithm for full-range imaging by suppression of the complex conjugate artifact in phase-shifting Fourier domain optical coherence tomography. This technique utilizes the projection of multiple phase-shifted interferograms onto an orthogonal basis set to reconstruct the complex interferogram. Full-range imaging with >30 dB suppression of the symmetric artifact is demonstrated using a 3 x 3 fiber coupler swept source OCT system, providing a depth range of 6.6mm with -8 dB roll-off in sensitivity at the depth boundaries relative to DC. Real-time display of full-range images of the anterior segment of the human eye acquired in vivo at a line rate of 6.67 kHz are presented.[Abstract] [Full Text] [Related] [New Search]