These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Impact of electrode insertion depth on intracochlear trauma.
    Author: Adunka O, Kiefer J.
    Journal: Otolaryngol Head Neck Surg; 2006 Sep; 135(3):374-82. PubMed ID: 16949967.
    Abstract:
    OBJECTIVE: To assess the effect of cochlear implant (CI) insertion depth and surgical technique on intracochlear trauma. STUDY DESIGN AND SETTING: Twenty-one fresh human temporal bones were implanted with CI electrodes and underwent histologic processing and evaluation. Specimens were grouped into 3 categories: 1) soft implantation technique and standard electrode; 2) soft implantation technique and flexible prototype array; 3) forceful implantations and standard electrode. Based on the grading system (1 to 4), 2 numeric values were calculated indicating the overall severity of cochlear damage (trauma indices). RESULTS: Mean trauma index values were 13.8, 36.3, and 59.2 for group 1, 2, and 3, respectively. Differences in cochlear trauma (trauma index) were nonsignificant between specimens in groups 1 and 2 but were significant between groups 1 and 3. CONCLUSION: This study gives evidence that intracochlear trauma increases with deep insertions. Thus, in cases where cochlear integrity might be important, limited insertions should be achieved.
    [Abstract] [Full Text] [Related] [New Search]