These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Single-detector micro-electro-mechanical scanning grating spectrometer.
    Author: Kraft M, Kenda A, Frank A, Scherf W, Heberer A, Sandner T, Schenk H, Zimmer F.
    Journal: Anal Bioanal Chem; 2006 Nov; 386(5):1259-66. PubMed ID: 16964475.
    Abstract:
    A compact, robust grating spectrometer based on an optimised micro-electro-mechanical grating mirror component has been developed, built, and characterised. The application of an oscillating reflection grating micro-mirror component as scanning dispersive element in a modified Czerny-Turner monochromator layout enables the design of compact grating spectrometers capable of acquiring full spectra using a single detector element. Designed for a wavelength range between 1200 and 1900 nm, the spectrometer features a spectral resolution of 10 nm with wavelength stability better than +/-0.5 nm. One-hundred scan spectra can be acquired in less than one second, or spectral changes can be monitored at time a resolution of less than 10 ms. In combination with a fibre-optic interface and a typical weight of less than 1 kg, this makes this novel type of fully portable micro-electro-mechanical near-IR scanning spectrometer an interesting alternative to existing spectrometers and opens a range of new applications, in particular the detection of major and minor components in the near-IR.
    [Abstract] [Full Text] [Related] [New Search]