These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Linear optical coherence tomography system with extended measurement range.
    Author: Koch P, Hellemanns V, Hüttmann G.
    Journal: Opt Lett; 2006 Oct 01; 31(19):2882-4. PubMed ID: 16969410.
    Abstract:
    Optical coherence tomography (OCT) sensors traditionally use scanning optical delay lines with moving parts and a single detector. OCT systems with a linear detector array (linear OCT or L-OCT) are simple and robust, but a detector with approximately 10,000 pixels is needed for an imaging depth of 2mm, which is necessary for most biomedical applications. We present a new optical setup for L-OCT with an increased measurement range. An additional grating performs a reduction of the spatial frequencies of the fringe pattern on the detector without loss in the signal-to-noise ratio, so the signal can be sampled with a minimal number of pixels. The theory for this approach is addressed and the first measurements are presented.
    [Abstract] [Full Text] [Related] [New Search]