Search MEDLINE/PubMed


  • Title: Application of the EMIT 2000 Tacrolimus assay on the Abbott Architect c8000 high volume clinical chemistry analyzer.
    Author: Boer K, Deufel T, Schmidt D, Streck S, Kiehntopf M.
    Journal: Clin Biochem; 2006 Oct; 39(10):1041-3. PubMed ID: 16979616.
    Abstract:
    OBJECTIVES: Evaluation of the performance of the EMIT 2000 Tacrolimus assay on the Abbott Architect c8000 analyzer. DESIGN AND METHODS: Imprecision studies were performed and patient samples were assayed by EMIT assay and by LC-MS/MS. RESULTS: Limit of quantification was established at 2.8 microg/L. A positive bias of 17.5% between results measured on EMIT and on LC-MS/MS was detected. CONCLUSIONS: EMIT 2000 Tacrolimus assay is suitable for automated analyses of Tacrolimus on the Architect c8000.