These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Determination of dosimetrical quantities used in microbeam radiation therapy (MRT) with Monte Carlo simulations.
    Author: Siegbahn EA, Stepanek J, Bräuer-Krisch E, Bravin A.
    Journal: Med Phys; 2006 Sep; 33(9):3248-59. PubMed ID: 17022219.
    Abstract:
    Microbeam radiation therapy (MRT) is being performed by using an array of narrow rectangular x-ray beams (typical beam sizes 25 microm X 1 cm), positioned close to each other (typically 200 microm separation), to irradiate a target tissue. The ratio of peak-to-valley doses (PVDR's) in the composite dose distribution has been found to be strongly correlated with the normal tissue tolerance and the therapeutic effect of MRT. In this work a Monte Carlo (MC) study of the depth- and lateral-dose profiles in water for single x-ray microbeams of different shapes and energies has been performed with the MC code PENELOPE. The contributions to the dose deposition from different interaction types have been determined at different distances from the center of the microbeam. The dependence of the peak dose, in a water phantom, on the microbeam field size used in the preclinical trials, has been demonstrated. Composite dose distributions for an array of microbeams were obtained using superposition algorithms and PVDR's were determined and compared with literature results obtained with other Monte Carlo codes. The dependence of the PVDR's on microbeam width, x-ray energy used, and on the separation between adjacent microbeams has been studied in detail.
    [Abstract] [Full Text] [Related] [New Search]