These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Correction of eddy-current distortions in diffusion tensor images using the known directions and strengths of diffusion gradients.
    Author: Zhuang J, Hrabe J, Kangarlu A, Xu D, Bansal R, Branch CA, Peterson BS.
    Journal: J Magn Reson Imaging; 2006 Nov; 24(5):1188-93. PubMed ID: 17024663.
    Abstract:
    PURPOSE: To correct eddy-current artifacts in diffusion tensor (DT) images without the need to obtain auxiliary scans for the sole purpose of correction. MATERIALS AND METHODS: DT images are susceptible to distortions caused by eddy currents induced by large diffusion gradients. We propose a new postacquisition correction algorithm that does not require any auxiliary reference scans. It also avoids the problematic procedure of cross-correlating images with significantly different contrasts. A linear model is used to describe the dependence of distortion parameters (translation, scaling, and shear) on the diffusion gradients. The model is solved numerically to provide an individual correction for every diffusion-weighted (DW) image. RESULTS: The assumptions of the linear model were successfully verified in a series of experiments on a silicon oil phantom. The correction obtained for this phantom was compared with correction obtained by a previously published method. The algorithm was then shown to markedly reduce eddy-current distortions in DT images from human subjects. CONCLUSION: The proposed algorithm can accurately correct eddy-current artifacts in DT images. Its principal advantages are that only images with comparable signals and contrasts are cross-correlated, and no additional scans are required.
    [Abstract] [Full Text] [Related] [New Search]