These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Appropriate result from an inappropriate ICD shock.
    Author: Davis DR, Gollob MH, Green MS, Lemery R, Tang AS, Birnie DH.
    Journal: Pacing Clin Electrophysiol; 2006 Oct; 29(10):1183-4. PubMed ID: 17038153.
    Abstract:
    A case of inappropriate detection of an ungrounded external AC current by an implantable cardioverter defibrillator (ICD) is reported. The resultant ICD shock dislodged the patient from the electrical source, thus preventing injury or electrocution.
    [Abstract] [Full Text] [Related] [New Search]