These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Snare-assisted anterograde balloon mitral and aortic valvotomy using Inoue balloon catheter.
    Author: Krishnan MN, Syamkumar MD, Sajeev CG, Venugopal K, Johnson F, Vinaykumar D, Velayudhan CC, Jayakumar TG.
    Journal: Int J Cardiol; 2007 Jan 02; 114(1):e9-11. PubMed ID: 17049634.
    Abstract:
    We performed concurrent antegrade mitral and aortic valvotomy using Inoue dilatation catheter in 3 cases of combined rheumatic mitral and aortic stenosis. Following mitral valvotomy by standard procedure, aortic valve was crossed with the help of a floatation catheter. Stiff long length guide wire was fixed in descending aorta using a snare. Inoue catheter was threaded over the wire across the aortic valve and aortic valvotomy completed. Mitral valve area increased from mean 1 cm2 to 2 cm2; aortic gradient dropped from mean of 97 mm to 36 mm. Concurrent anterograde balloon mitral and aortic valvotomy may be effective and safe.
    [Abstract] [Full Text] [Related] [New Search]