These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: The development of a real-time biosensor for the detection of trace levels of trinitrotoluene (TNT) in aquatic environments. Author: Bromage ES, Lackie T, Unger MA, Ye J, Kaattari SL. Journal: Biosens Bioelectron; 2007 May 15; 22(11):2532-8. PubMed ID: 17088054. Abstract: This paper describes the development of a highly sensitive TNT immunosensor consisting of a highly specific monoclonal antibody coupled with a prototype fluorescence-based detector system (KinExA Inline Biosensor, Sapidyne Instrument Inc). The antibody developed possesses a high affinity for TNT (association constant, aK 8.2) with minimal cross reactivity with other compounds such as tetryl, 2,4-dinitrotoluene and 2-amino-4,6-dinitrotoluene. This system provides sample assessment within 160 s from source acquisition and possesses sensitivity for TNT of 0.05 microg/L in ground water. The sensor can be regenerated in 8 min, allows a minimum of 40 repeated readings, and has a standard error of 0.1-0.4% between repeat readings. The fluidics and software allow samples to be obtained from up to eight different sources allowing the user to examine the stratification of the pollutant in the water column. We believe that this immunosensor can be used to rapidly assess trace levels of TNT in environmental water samples.[Abstract] [Full Text] [Related] [New Search]