These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Prospective evaluation of double RBC collection using three different apheresis systems. Author: Picker SM, Radojska SM, Gathof BS. Journal: Transfus Apher Sci; 2006 Dec; 35(3):197-205. PubMed ID: 17097350. Abstract: BACKGROUND: Automated component collection systems offer the possibility of increasing blood supply and improving transfusion safety. DESIGN: 30 blood donors were randomly assigned to double RBC collection with either the Baxter Alyx (AX), the Haemonetics MCS Plus (MCS+), or the Gambro Trima Accel (TA). Procedures were prospectively evaluated focussing on yield, time, efficiency, citrate donor load, and in vitro quality. RESULTS: All units showed sufficient in vitro quality throughout 42 days of storage and complied with international requirements. Donor reactions were limited to mild citrate reactions. AX was the fastest and most efficient system* * (* *p approximately 0.001) attaining the highest yield* * from similar amounts of whole blood. The drawbacks were a higher RBC loss* (*p < 0.05) and accelerated citrate infusion* *. Due to lower collection rates* * * (* * *p < 0.001), MCS+ was slower than TA* * * but compensated with lower citrate load * * *. CONCLUSION: Double RBC apheresis was performed safely and efficiently with all three instruments. AX had advantages for most parameters evaluated.[Abstract] [Full Text] [Related] [New Search]