These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Hearing threshold assessment post grommet insertion. Is it reliable?
    Author: Griffiths H, James D, Davis R, Hartland S, Molony N.
    Journal: J Laryngol Otol; 2007 May; 121(5):431-4. PubMed ID: 17164021.
    Abstract:
    We aimed to investigate whether auditory brainstem response (ABR), assessed immediately after grommet insertion, was an accurate measure of hearing threshold. Ten children were recruited (20 ears). Patients underwent transient evoked otoacoustic emission (TEOAE) and ABR tests before and after grommet insertion under general anaesthetic. A third assessment was conducted at a six-week review, without anaesthetic or sedation. The hearing threshold in dB and the presence of TEOAE were recorded by senior audiologists. The results of each investigation were compared. At review, only 15 ABR and 17 TEOAE tests, of a possible 20, were recorded. Transient evoked otoacoustic emissions were recorded in 30 per cent of ears pre-operatively, 20 per cent of ears post-operatively and 82 per cent of ears at six-week review. The average ABR thresholds were 23.75 dB (range 10-45 dB) pre-operatively, 27.25 (range 15-40 dB) post-operatively and 15 dB (range 5-25 dB) at six-week review. Comparing pre- and post-operative assessment, decreased thresholds were seen in six ears and elevated thresholds in 11. Comparing the six-week review ABR with the post-operative ABR, 13 ears had decreased thresholds, by as much as 20 dB. Thresholds of 35 dB or greater were measured in six ears immediately after grommet insertion (maximum 40 dB). Threshold shifts of 15 dB or greater was observed in seven ears. The intra-subject variation was shown to be significant (p<0.001) when repeated measures analysis of variance was applied. We conclude that there is evidence that ABR can be inaccurate after grommet insertion, overestimating hearing thresholds. We believe that the mechanism by which the error is realised is temporary threshold shift.
    [Abstract] [Full Text] [Related] [New Search]