These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: High-performance liquid chromatography-electrospray ionization mass spectrometry determination of mitiglinide in human plasma and its pharmacokinetics.
    Author: Liang J, Tian Y, Zhang Z, Feng S, Zhao Y, Mao G.
    Journal: J Mass Spectrom; 2007 Feb; 42(2):171-7. PubMed ID: 17173338.
    Abstract:
    A selective and sensitive method employing high-performance liquid chromatography-electrospray ionization mass spectrometry was developed and validated for the determination of mitiglinide in human plasma.With gliclazide as the internal standard, mitiglinide was extracted from plasma with n-hexane: = 80 : 20 (v/v). The organic layer was evaporated and the residue was redissolved in methanol: water (10 mM CH3COONH4, pH = 3.0) = 65 : 35 (v/v). An aliquot of 10 microl was chromatographically analyzed on a prepacked Shimadzu VP-ODS (5 microm, 150 x 2.0 mm i.d.) using the mobile phase comprising methanol: water (10 mM CH3COONH4) = 65 : 35 (v/v) by means of selected-ion monitoring mode mass spectrometry. Standard curves were linear (r2 = 0.9972) over the concentration range of 2.84-11 300 pmol/ml and had good accuracy and precision. The within- and between-batch precisions of the method were within 15% of standard deviation. The lower limit of detection was 1.42 pmol/ml. The validated HPLC/ESI-MS method has been successfully applied in the pharmacokinetics of mitiglinide in 12 healthy Chinese volunteers.
    [Abstract] [Full Text] [Related] [New Search]