  • Title: [Results of usage of temporary prosthetics in cases of combat injuries of extremities].
    Author: Samokhvalov IM, Zavrazhanov AA, Kornilov EA.
    Journal: Voen Med Zh; 2006 Sep; 327(9):29-33, 96. PubMed ID: 17180855.
    Abstract:
    The comparative analyzes of the results of usage of temporary prosthetics (TP) of injured main arteries during the war in Afghanistan (1979-1989) and counter-terrorist operations in the Northern Caucasus (1994-1996, 1999-2002) has been carried out. The development of new models of temporary prosthetics of vessels has been experimentally substantiated. In Afghanistan the two-staged usage of TP was made along with the amputation of extremities in 35,1% of cases and in the Northern Caucasus - in 15,0%. The reason for that was the considerable reduction of terms of evacuation of wounded. The average terms of presence of TP in the arteries in Afghanistan were 26,2+/-4,0 hours and in the Northern Caucasus - 18,1+/-3,4 hours. During a series of experiments on comparative testing of efficiency of three types of TP it was determined that prosthetics made of armored silicone (10,1+/-0,6 hours)and, especially made of enlarged micropored polytetraftoraethylen (11,5 +/-0,6 hours) are thrombosed slower comparing to polyclorvinil tube from the blood transfusion system (7,9+/-0,7 hours).