These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: [Results of usage of temporary prosthetics in cases of combat injuries of extremities]. Author: Samokhvalov IM, Zavrazhanov AA, Kornilov EA. Journal: Voen Med Zh; 2006 Sep; 327(9):29-33, 96. PubMed ID: 17180855. Abstract: The comparative analyzes of the results of usage of temporary prosthetics (TP) of injured main arteries during the war in Afghanistan (1979-1989) and counter-terrorist operations in the Northern Caucasus (1994-1996, 1999-2002) has been carried out. The development of new models of temporary prosthetics of vessels has been experimentally substantiated. In Afghanistan the two-staged usage of TP was made along with the amputation of extremities in 35,1% of cases and in the Northern Caucasus - in 15,0%. The reason for that was the considerable reduction of terms of evacuation of wounded. The average terms of presence of TP in the arteries in Afghanistan were 26,2+/-4,0 hours and in the Northern Caucasus - 18,1+/-3,4 hours. During a series of experiments on comparative testing of efficiency of three types of TP it was determined that prosthetics made of armored silicone (10,1+/-0,6 hours)and, especially made of enlarged micropored polytetraftoraethylen (11,5 +/-0,6 hours) are thrombosed slower comparing to polyclorvinil tube from the blood transfusion system (7,9+/-0,7 hours).[Abstract] [Full Text] [Related] [New Search]