  • Title: An integrated system for multichannel neuronal recording with spike/LFP separation, integrated A/D conversion and threshold detection.
    Author: Perelman Y, Ginosar R.
    Journal: IEEE Trans Biomed Eng; 2007 Jan; 54(1):130-7. PubMed ID: 17260864.
    Abstract:
    A mixed-signal front-end processor for multichannel neuronal recording is described. It receives 12 differential-input channels of implanted recording electrodes. A programmable cutoff High Pass Filter (HPF) blocks dc and low-frequency input drift at about 1 Hz. The signals are band-split at about 200 Hz to low-frequency Local Field Potential (LFP) and high-frequency spike data (SPK), which is band limited by a programmable-cutoff LPF, in a range of 8-13 kHz. Amplifier offsets are compensated by 5-bit calibration digital-to-analog converters (DACs). The SPK and LFP channels provide variable amplification rates of up to 5000 and 500, respectively. The analog signals are converted into 10-bit digital form, and streamed out over a serial digital bus at up to 8 Mbps. A threshold filter suppresses inactive portions of the signal and emits only spike segments of programmable length. A prototype has been fabricated on a 0.35-microm CMOS process and tested successfully, demonstrating a 3-microV noise level. Special interface system incorporating an embedded CPU core in a programmable logic device accompanied by real-time software has been developed to allow connectivity to a computer host.