These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Analysis of carvedilol in human plasma using hydrophilic interaction liquid chromatography with tandem mass spectrometry. Author: Jeong DW, Kim YH, Ji HY, Youn YS, Lee KC, Lee HS. Journal: J Pharm Biomed Anal; 2007 Jun 28; 44(2):547-52. PubMed ID: 17270380. Abstract: A rapid, sensitive and selective method for the determination of carvedilol in human plasma was developed using hydrophilic interaction liquid chromatography with tandem mass spectrometry (HILIC-MS/MS). Carvedilol and cisapride (internal standard) were extracted from human plasma with methyl tert-butyl ether at basic pH and analyzed on an Atlantis HILIC Silica column with the mobile phase of acetonitrile-ammonium formate (50 mM, pH 4.5) (90:10, v/v). The analytes were detected using an electrospray ionization tandem mass spectrometry in the multiple-reaction-monitoring mode. The standard curve was linear (r=0.9998) over the concentration range of 0.1-200 ng/ml. The lower limit of quantification for carvedilol was 0.1 ng/ml using 50 microl plasma sample. The coefficient of variation and relative error for intra- and inter-assay at four QC levels were 1.6-4.5% and -6.4 to 4.8%, respectively. The absolute and relative matrix effect for carvedilol and cisapride were practically absent. The extraction recoveries of carvedilol and cisapride were 81.6 and 85.2%, respectively. This method was successfully applied to the bioequivalence study of carvedilol in humans.[Abstract] [Full Text] [Related] [New Search]