These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Development and validation of a method for quantitative determination of valsartan in human plasma by liquid chromatography-tandem mass spectrometry. Author: Koseki N, Kawashita H, Hara H, Niina M, Tanaka M, Kawai R, Nagae Y, Masuda N. Journal: J Pharm Biomed Anal; 2007 Apr 11; 43(5):1769-74. PubMed ID: 17289324. Abstract: A sensitive liquid chromatography-tandem mass spectrometry (LC-MS/MS) method for the determination of valsartan in human plasma was developed and validated. A 0.5 ml aliquot was extracted using solid-phase extraction in an Empore high performance extraction disk plate, universal resin 96-well format. The estimated calibration range of the method was 2-2000 ng/ml. The method was fully validated with intra-day mean accuracy and precision of 94.8-107% and 2.19-5.40% and inter-day mean accuracy and precision of 93.5-105% and 1.87-5.67%, respectively. No significant loss of valsartan in processed samples was confirmed in processed samples for up to 24 h at 10 degrees C. Sample dilution up to 50-fold with blank human plasma provided acceptable analyses. No interference peaks or matrix effects were observed. No effect of QC sample location results was observed in a 96-well plate. This LC-MS/MS technique was found to improve quantitative determination of valsartan allowing its pharmacokinetic evaluation with clinically relevant doses.[Abstract] [Full Text] [Related] [New Search]