These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Electrical properties of retinal-electrode interface. Author: Shah S, Hines A, Zhou D, Greenberg RJ, Humayun MS, Weiland JD. Journal: J Neural Eng; 2007 Mar; 4(1):S24-9. PubMed ID: 17325413. Abstract: A critical element of a retinal prosthesis is the stimulating electrode array, which is placed in close proximity to the retina. It is via this retinal-electrode interface that a retinal prosthesis electrically stimulates nerve cells to produce the perception of light. The impedance load seen by the current driver consists of the tissue resistance and the complex electrode impedance. The results in this paper show that the tissue resistance of the retina is significantly greater than that of the vitreous humor in the eye. Circuit models of the electrode-retina interface are used to parameterize the different contributors to the overall impedance.[Abstract] [Full Text] [Related] [New Search]