These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Fluorescence dilution technique for measurement of cardiac output and circulating blood volume in healthy human subjects. Author: Maarek JM, Holschneider DP, Rubinstein EH. Journal: Anesthesiology; 2007 Mar; 106(3):491-8. PubMed ID: 17325507. Abstract: BACKGROUND: The authors previously validated in an animal model a new indicator dilution technique for measuring cardiac output and circulating blood volume by recording transcutaneously the fluorescence of circulating indocyanine green with an optical probe placed on the skin surface. The current study compared fluorescence dilution recordings recorded from several locations on the human face in terms of signal intensity and stability and estimated the subjects' cardiac output and circulating blood volume from the recordings. METHODS: Fluorescence dilution traces were recorded transcutaneously in six healthy human volunteers after rapid intravenous injection of 1 mg indocyanine green. Three placements of the optical probe were tested: the nose ala, the ear lobe, and the temple area. In three subjects, the recordings were calibrated in terms of circulating indocyanine green concentration to estimate cardiac output and circulating blood volume. RESULTS: Fluorescence dilution traces had the same duration for the three locations, but the recordings obtained from the nose ala and the ear lobe were twice as intense as those from the temple. The fluorescence intensity at each site was linearly related to the local laser Doppler perfusion index. The coefficient of variation for the area under the first pass curve (inversely proportional to cardiac output) was approximately 6% for triplicate measurements at the same location. Cardiac output and circulating blood volume derived from the fluorescence recordings were in the normal range. CONCLUSIONS: The study demonstrates that intense and reproducible fluorescence dilution signals can be measured transcutaneously in healthy humans and could potentially be used to measure cardiac output and circulating blood volume minimally invasively.[Abstract] [Full Text] [Related] [New Search]