These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: A comparison of transcutaneous end-tidal and arterial measurements of carbon dioxide during general anaesthesia. Author: Reid CW, Martineau RJ, Miller DR, Hull KA, Baines J, Sullivan PJ. Journal: Can J Anaesth; 1992 Jan; 39(1):31-6. PubMed ID: 1733530. Abstract: A randomized, prospective study was performed to evaluate the accuracy of a new transcutaneous carbon dioxide (CO2) monitor (Fastrac) during general anaesthesia. Twenty-two adult patients undergoing elective surgery were subjected to three different levels of minute ventilation by varying their respiratory rates in a randomized cross-over design. Simultaneous measurements of transcutaneous CO2 (PTCCO2) and arterial CO2 (PaCO2) were obtained at three levels of minute ventilation (low, medium and high). End-tidal CO2 (PETCO2) values were also recorded from a mass spectrometer (SARA) at each time period. A total of 66 data sets with PaCO2 ranging from 28-62 mmHg were analyzed. The PTCCO2 values demonstrated a high degree of correlation with PaCO2 over the range of minute ventilation (y = 0.904x + 6.36, r = 0.92, P less than 0.001). The PETCO2 measurement also demonstrated a generally good correlation with PaCO2 (y = 0.62x + 9.21, r = 0.89, and P less than 0.01). However, the PETCO2-PaCO2 gradients (mean 7.0 +/- 3.1 mmHg) were greater than the PTCCO2-PaCO2 gradients (mean 2.3 +/- 2.4 mmHg) at all three levels of minute ventilation (P less than 0.05). These differences were greatest when PaCO2 was in the high range (48-60 mmHg). We conclude that the new Fastrac CO2 monitor is accurate for monitoring carbon dioxide levels during general anaesthesia. The new transcutaneous devices provide an effective method for non-invasive monitoring of CO2 in situations where continuous, precise control of CO2 levels is desired.[Abstract] [Full Text] [Related] [New Search]