These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Elimination of matrix effects in electrothermal atomic absorption spectrophotometric determinations of bismuth in serum and urine. Author: Dean S, Tscherwonyi PJ, Riley WJ. Journal: Clin Chem; 1992 Jan; 38(1):119-22. PubMed ID: 1733583. Abstract: A sensitive and precise electrothermal atomic absorption spectrophotometric method for determining bismuth concentration is described. Protein precipitation and the use of a palladium modifier reduce the problems of foaming and permit the use of a higher ashing temperature. The detection limit of the assay is 0.9 nmol/L. Total CVs (intra- and interassay) for serum ranged from 3.5% to 15.1% and for urine from 4.8% to 14.5% at concentrations of 60.0 and 6.0 nmol/L, respectively. Analytical recoveries of bismuth added to serum and urine were 102% and 103% over the same range. The method is robust and reproducible and can be accurately calibrated with aqueous standards.[Abstract] [Full Text] [Related] [New Search]