These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Use of the kinetic plot method to analyze commercial high-temperature liquid chromatography systems. II. Practically constrained performance comparison.
    Author: Clicq D, Heinisch S, Rocca JL, Cabooter D, Gzil P, Desmet G.
    Journal: J Chromatogr A; 2007 Apr 06; 1146(2):193-201. PubMed ID: 17336314.
    Abstract:
    Using a set of experimentally determined plate height data obtained on three commercial high-temperature HPLC supports, and evaluating their isocratic separation speed potential under the application of a set of instrumental constraints, a qualitative map of the practically achievable critical pair separation speed potential of high-temperature HPLC has been established. The obtained data show that the gain in separation speed is more strongly affected by the instrumental limitations in the high-temperature range than it is for the low temperatures. For the presently considered case of alkylbenzene separations, the potential gain in analysis time that can be obtained by going from T=30 to 120 degrees C in the presence of a typical set of instrumental limitations nevertheless remains of the order of a factor of 2-4. The study also shows that improvements on the instrumentation side (increased detector frequency, pumping flow rate, smaller extra-column volumes, ...) are indispensable to fully benefit from the high temperature advantages for all separations requiring less than 10,000 effective theoretical plates.
    [Abstract] [Full Text] [Related] [New Search]