Title: Use of the kinetic plot method to analyze commercial high-temperature liquid chromatography systems. II. Practically constrained performance comparison. Author: Clicq D, Heinisch S, Rocca JL, Cabooter D, Gzil P, Desmet G. Journal: J Chromatogr A; 2007 Apr 06; 1146(2):193-201. PubMed ID: 17336314. Abstract: Using a set of experimentally determined plate height data obtained on three commercial high-temperature HPLC supports, and evaluating their isocratic separation speed potential under the application of a set of instrumental constraints, a qualitative map of the practically achievable critical pair separation speed potential of high-temperature HPLC has been established. The obtained data show that the gain in separation speed is more strongly affected by the instrumental limitations in the high-temperature range than it is for the low temperatures. For the presently considered case of alkylbenzene separations, the potential gain in analysis time that can be obtained by going from T=30 to 120 degrees C in the presence of a typical set of instrumental limitations nevertheless remains of the order of a factor of 2-4. The study also shows that improvements on the instrumentation side (increased detector frequency, pumping flow rate, smaller extra-column volumes, ...) are indispensable to fully benefit from the high temperature advantages for all separations requiring less than 10,000 effective theoretical plates.[Abstract] [Full Text] [Related] [New Search]