These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: A simple competitive RT-PCR assay for quantitation of HIV-1 subtype B and non-B RNA in plasma. Author: Hamatake M, Nishizawa M, Yamamoto N, Kato S, Sugiura W. Journal: J Virol Methods; 2007 Jun; 142(1-2):113-7. PubMed ID: 17367872. Abstract: An easy, inexpensive competitive RT-PCR assay for HIV-1 RNA quantitation was constructed. A 138-bp sequence in the HIV-1 gag p24 region was selected as the target and co-amplified with competitor RNA containing an internal 44-bp deletion. Quantitation of serial dilutions of control RNA samples prepared from the LAI isolate demonstrated a good linearity (R(2)=0.991) within the range between 10 and 250 copies/sample. The detection limit of the assay was determined to be 3.8 copies/sample by Probit analysis and corresponded to 110 copies/ml in plasma. The intra-assay CV value was 9.1%, and the inter-assay value was 25.9%. Both were comparable to those obtained with commercially available HIV-1 RNA quantitation kits. The correlation efficient for the results obtained in 47 plasma samples from HIV-1-infected individuals (subtype A in 1, subtype B in 25, subtype C in 4, subtype F in 1, and CRF01 AE in 16) with the competitive RT-PCR and Cobas Amplicor HIV-1 Monitor test v1.5 was 0.956 for subtype B and 0.947 for subtype non-B. The assay devised is a good alternative for monitoring antiretroviral therapy in resource-poor countries.[Abstract] [Full Text] [Related] [New Search]