Title: Cloud point extraction of trace metals from seawater and determination by electrothermal atomic absorption spectrometry with iridium permanent modifier. Author: Meeravali NN, Reddy MA, Kumar SJ. Journal: Anal Sci; 2007 Mar; 23(3):351-6. PubMed ID: 17372381. Abstract: A simple method is described for preconcentration and separation of trace metals such as Ag, Co, Cr, Cu, Fe, Mn, Ni and Pb simultaneously from seawater using a cloud point extraction (CPE) procedure. Triton X-114 nonionic surfactant and ammonium pyrrolidine dithiocarbamate (APDC) have been used as an extraction medium and a chelating extractant, respectively. The amounts of Triton X-114 and APDC and the pH value necessary for extraction were carefully optimized. The preconcentration factor of about 200 is achieved for all the studied metals. Electrothermal atomic absorption spectrometry (ETAAS) with an Ir coated graphite tube as permanent chemical modifier has been used for determination. The limits of detection of Ag, Co, Cr, Cu, Fe, Mn, Ni and Pb were 0.003, 0.008, 0.003, 0.006, 0.015, 0.002, 0.009 and 0.01 ng ml-1, respectively. Certified reference materials such as CASS-4 and NASS-5 (seawater) and NIST-1640 (natural water) have been used for validation of the new method. The relative standard deviation (%) obtained for all the metals are in the range 0.8 - 3.6% for natural water and 11-25% for seawater materials, except for Co in NASS-5 for which it was 50%.[Abstract] [Full Text] [Related] [New Search]