These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Raman sensitivity enhancement for aqueous absorbing sample using Teflon-AF 2400 liquid core optical fibre cell. Author: Tian Y, Zhang L, Zuo J, Li Z, Gao S, Lu G. Journal: Anal Chim Acta; 2007 Jan 02; 581(1):154-8. PubMed ID: 17386439. Abstract: The compatibility Teflon-AF 2400 liquid core optical fibre with resonance Raman spectroscopy (RRS-LCOF) was used to detect aqueous biomolecules. The maximum sensitivity enhancement factor for concentrations greater than the detection limit in a conventional cell was 10, and detection limit reduction of about 1000-fold have been achieved for the measurement of aqueous absorbing sample using Teflon-AF 2400 fibre Raman cell compared to the conventional cell. We were able to collect spectra of 2.5 x 10(-9) and 2.5 x 10(-10) M aqueous beta-carotene using 16.2 mW of laser power and 10s integration time. This volume of a 2.5 x 10(-10) M aqueous solution corresponds to only 1.5 fmol or 830 fg of beta-carotene. The results of this preliminary study indicate that RRS-LCOF has potential in bioanalytical and biomedical applications.[Abstract] [Full Text] [Related] [New Search]