These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Potentiometric evaluation of calix[4]arene anion receptors in membrane electrodes: phosphate detection.
    Author: Kivlehan F, Mace WJ, Moynihan HA, Arrigan DW.
    Journal: Anal Chim Acta; 2007 Feb 28; 585(1):154-60. PubMed ID: 17386660.
    Abstract:
    Ion-selective membrane electrodes doped with the urea- or thiourea-functionalised calix[4]arenes, 5,11,17,23-tetra-tert-butyl-25,27-bis[[4-N'-(phenylureido)butyl]oxy]-26,28-dipropoxy calix[4]arene (I) and 5,11,17,23-tetra-tert-butyl-25,27-bis[[4-(N'-phenylthioureido)-butyl]oxy]-26,28-dipropoxy calix[4]arene (II), were evaluated for anion sensing. Potentiometric results show that these calixarene ionophore-based membrane electrodes exhibit a good sensitivity to aqueous solutions of the monohydrogen orthophosphate species HPO(4)(2-) in the concentration range 5.0 x 10(-5) to 1.0 x 10(-1)M, with near-Nernstian response slopes of -33.0 and -28.0 mV dec(-1) for ionophores I and II, respectively. Selectivity coefficient values for monohydrogen orthophosphate over a range of common anions were determined by the fixed interference and matched potential methods and indicated that these membrane electrodes exhibit a good selectivity for HPO(4)(2-) with respect to the other anions, including sulfate and nitrate.
    [Abstract] [Full Text] [Related] [New Search]