These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Rates of and reasons for failure of commercial human immunodeficiency virus type 1 viral load assays in Brazil. Author: Drexler JF, de Souza Luna LK, Pedroso C, Pedral-Sampaio DB, Queiroz AT, Brites C, Netto EM, Drosten C. Journal: J Clin Microbiol; 2007 Jun; 45(6):2061-3. PubMed ID: 17392448. Abstract: We examined failures of commercial human immunodeficiency virus type 1 (HIV-1) viral load assays of 1,195 plasma samples from Brazilian patients. Assay failure was assumed for samples in which the virus was undetectable by commercial assay but which tested positive by real-time reverse transcription-PCR of the HIV-1 long terminal repeat (LTR) region or if the viral load differed by >2 log10 from that determined by LTR assay. Failure rates for Bayer Versant bDNA 3.0, Roche Amplicor Monitor v1.5, and bioMerieux NucliSens QT were 0.68, 0.47, and 4.33%, respectively. NucliSens may be inadequate for use in Brazil.[Abstract] [Full Text] [Related] [New Search]