These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Measurement of cardiac output--transtracheal Doppler versus thermodilution. Author: Froese N, Friesen R. Journal: Can J Anaesth; 1991 Oct; 38(7):931-4. PubMed ID: 1742833. Abstract: The ABCOM 1 transtracheal Doppler (TTD) has been developed as a non-invasive cardiac output monitor. With this device, cardiac output is continuously calculated from ascending aortic blood flow velocity and aortic diameter obtained via an ultrasound transducer incorporated into the tip of an endotracheal tube. We evaluated the clinical use of the ABCOM 1 monitor and compared cardiac outputs obtained using the TTD system with simultaneous thermodilution (TD) measurements. We found the operation of the ABCOM 1 monitor to be difficult and time-consuming. In our operating rooms, acceptable Doppler signal quality was difficult to obtain. There was no correlation between 36 simultaneously obtained TTD and TD cardiac output measurements. The average difference between measurement techniques and the limits of agreement were unacceptably large (mean difference = 3.04 L.min-1, mean +/- 2 SD = -6.04 to 12.48 L.min-1). Separately analyzing only those measurements during which Doppler signal quality was adequate did not improve agreement between TTD and TD measurements. On the basis of these findings, TTD cannot be recommended as a clinical cardiac output measurement technique.[Abstract] [Full Text] [Related] [New Search]