These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Ultratrace determination of inorganic selenium without signal calibration. Author: Hazelton SG, Pierce DT. Journal: Anal Chem; 2007 Jun 15; 79(12):4558-63. PubMed ID: 17508714. Abstract: To more readily evaluate the complex biogeochemistry of selenium, a flow-through electrochemical method was developed that can accurately determine Se(IV) concentrations in aqueous samples to part-per-trillion levels without signal calibration. Stripping methods were used in conjunction with a high-efficiency, flow-through cell. The cell was designed with a novel gold working electrode that was separated from a porous counter electrode by a Nafion membrane. Because this design permitted exhaustive deposition of selenium from the sample stream as well as efficient coulometric stripping, determinations obeyed Faraday's law over a reasonably wide range of operating conditions. The method had a minimum quantitation limit of approximately 8 ng and a maximum limit of 800 ng for Se(IV). It was reliable for sample volumes as small as 0.5 mL and as high as 20 mL, thereby allowing determinations from part-per-million to just below part-per-billion levels. Interferences from Cu(II) and arsenate were evident, but only when these species were present at concentrations exceeding 10 mg.L-1. Overall, the method had a performance comparable to hydride-generation atomic absorption spectrometry but with less cumbersome equipment and freedom from calibration.[Abstract] [Full Text] [Related] [New Search]