These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Neutron fluence rate measurement using prompt gamma rays. Author: Vega-Carrillo HR, Manzanares-Acuña E, Hernández-Dávila VM, Chacón-Ruíz A, Gallego E, Lorente A. Journal: Radiat Prot Dosimetry; 2007; 126(1-4):265-8. PubMed ID: 17513856. Abstract: A gamma ray spectrometer, with a 3('') Ø X 3('') NaI(Tl) detector, with a moderator sphere has been utilised to measure the neutron fluence rate, with this value the H(10) was estimated. When a neutron is captured by the hydrogen-based moderator, a 2.22 MeV prompt gamma ray is produced. In a multichannel analyser the net area under the 2.22 MeV photopeak is proportional to the total neutron fluence rate. The features of this system were determined by a Monte Carlo study that includes 3-, 5- and 10-inches diameter, water and polyethylene moderators and a (239)Pu-Be source. The prompt gamma response was extended to monoenergetic neutron sources. To verify the response, a (239)Pu-Be source in combination with a 10('') polyethylene sphere having a gamma-ray spectrometer with NaI(Tl) was utilised to estimate the neutron fluence rate and the H(10). These results were compared with neutron fluence rate and H(10) obtained using a Bonner sphere spectrometer and with the H(10) measured using a neutron remmeter.[Abstract] [Full Text] [Related] [New Search]