These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Black-blood T2* technique for myocardial iron measurement in thalassemia. Author: He T, Gatehouse PD, Kirk P, Tanner MA, Smith GC, Keegan J, Mohiaddin RH, Pennell DJ, Firmin DN. Journal: J Magn Reson Imaging; 2007 Jun; 25(6):1205-9. PubMed ID: 17520740. Abstract: PURPOSE: To compare the effectiveness and reproducibility of a new black-blood sequence vs. a conventional bright-blood gradient-echo T2* sequence for myocardial iron overload measurement in thalassemia. MATERIALS AND METHODS: Twenty thalassemia patients were studied. Black-blood sequence images were acquired in diastole after a double inversion recovery (DIR) preparation pulse. Bright-blood sequence images were acquired in both early systole and late diastole. The data were randomized and the T2* analysis was performed blindly by two independent observers. RESULTS: The T2* values from the black-blood sequence were comparable to those of the conventional bright-blood sequence (25.7 +/- 12.9 msec vs. 26.4 +/- 14.2 msec in early systole, P = 0.44; and 25.2 +/- 13.1 msec in late diastole, P = 0.41). The coefficient of variation (CV) for black-blood image T2* analysis was 4.1% compared with 8.9% (early systole P = 0.03) and 7.8% (late diastole P = 0.05) for bright-blood image analysis. CONCLUSION: The black-blood T2* technique yields high-contrast myocardial images, provides clearly depicted myocardial borders, and avoids blood signal contamination of the myocardium while yielding improvements in interobserver variability.[Abstract] [Full Text] [Related] [New Search]