These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Hemodynamic simulation in a novel design for femoral bypass grafts.
    Author: Qiao A, Matsuzawa T.
    Journal: Biorheology; 2007; 44(2):103-24. PubMed ID: 17538202.
    Abstract:
    The effectiveness of femoral bypass grafts is correlated with the geometric configuration and hemodynamics of the bypass and the arteries. As an attempt to develop a new design for femoral bypass grafts, we present a novel geometric configuration for a symmetrically implanted 2-way bypass graft. In order to investigate how the symmetric 2-way bypass grafts affect the flow patterns through the anastomosis, physiologic blood flows in 1-way and 2-way models for a fully stenosed femoral bypass were simulated with the finite element method, and the hemodynamic factors in these models were studied. The temporal and spatial distributions of flow patterns and wall shear stresses in the vicinity of distal anastomosis during the cardiac cycle were analyzed. The results computed showed that the 2-way model has more preferable hemodynamics than the 1-way model in the distribution of flow patterns and wall shear stresses, and it may improve the flow conditions and decrease the probability of restenosis. However, the limitations of the 2-way bypass model may counteract the positive effects. More detailed hemodynamic studies are necessary to fully assess the viability of the 2-way bypass graft.
    [Abstract] [Full Text] [Related] [New Search]