These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Development of a neutron personal dose equivalent detector.
    Author: Tsujimura N, Yoshida T, Takada C, Momose T, Nunomiya T, Aoyama K.
    Journal: Radiat Prot Dosimetry; 2007; 126(1-4):261-4. PubMed ID: 17545659.
    Abstract:
    A new neutron-measuring instrument that is intended to measure a neutron personal dose equivalent, H(p)(10) was developed. This instrument is composed of two parts: (1) a conventional moderator-based neutron dose equivalent meter and (2) a neutron shield made of borated polyethylene, which covers a backward hemisphere to adjust the angular dependence. The whole design was determined on the basis of MCNP calculations so as to have response characteristics that would generally match both the energy and angular dependencies of H(p)(10). This new instrument will be a great help in assessing the reference values of neutron H(p)(10) during field testing of personal neutron dosemeters in workplaces and also in interpreting their readings.
    [Abstract] [Full Text] [Related] [New Search]