These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Quantitative determination of azithromycin in human plasma by ultra performance liquid chromatography-electrospray ionization mass spectrometry and its application in a pharmacokinetic study. Author: Chen L, Qin F, Ma Y, Li F. Journal: J Chromatogr B Analyt Technol Biomed Life Sci; 2007 Aug 15; 855(2):255-61. PubMed ID: 17574935. Abstract: A selective, rapid and sensitive ultra performance liquid chromatography-tandem mass spectrometry (UPLC-MS/MS) method was developed for the quantitative determination of azithromycin in human plasma and its application in a pharmacokinetic study. With roxithromycin as internal standard, sample pretreatment involved a one-step extraction with diethyl ether of 0.5 mL plasma. The analysis was carried out on an ACQUITY UPLC BEH C(18) column (50 mm x 2.1 mm, i.d., 1.7 microm) with gradient elution at flow rate of 0.35 mL/min. The mobile phase was 50 mM ammonium acetate and acetonitrile. The detection was performed on a triple-quadrupole tandem mass spectrometer by multiple reaction monitoring (MRM) mode via electrospray ionization (ESI). Linear calibration curves were obtained in the concentration range of 1-1000 ng/mL, with a lower limit of quantification of 1 ng/mL. The intra- and inter-day precision (RSD) values were below 15% and accuracy (RE) was -1.3% to 5.7% at all QC levels. The method was applicable to clinical pharmacokinetic study of azithromycin in healthy volunteers following oral administration.[Abstract] [Full Text] [Related] [New Search]