These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Spectrometer for lanthanides' K x-ray fluorescence.
    Author: Sakurai K, Mizusawa M, Terada Y.
    Journal: Rev Sci Instrum; 2007 Jun; 78(6):066108. PubMed ID: 17614652.
    Abstract:
    X-ray fluorescence analysis is a highly useful technique for determining the chemical composition of matter. The present article describes the successful development of a wavelength-dispersive x-ray fluorescence spectrometer for a fairly high-energy range, 30-60 keV, that can contribute to studying lanthanides' Kbeta spectra with high-energy resolution. By combining a new high-energy synchrotron light source and the present spectrometer, it has been demonstrated that the full width at half maximum for lanthanum's Kbeta(1) is 32 eV and that all the peaks in the spectra are fully resolved. This corresponds to an energy resolution EDeltaE of 1180, which is ten times better than a conventional system based on a Ge detector, which can detect only two peaks, Kbeta(1) and Kbeta(2), in seven peaks. The present spectrometer can open up a new field in x-ray spectrometry.
    [Abstract] [Full Text] [Related] [New Search]