Search MEDLINE/PubMed
Title: [Comparison of NucliSens HIV-1 QT and Amplicor HIV-1 monitor 1.5 in detecting HIV-1 viral load]. Author: Pan PL, Xu WY, Yao J, Tao XX, Pei LJ, Jiang Y. Journal: Zhonghua Shi Yan He Lin Chuang Bing Du Xue Za Zhi; 2007 Jun; 21(2):177-9. PubMed ID: 17653329. Abstract: OBJECTIVE: To compare the results of detecting HIV-1 load by using NucliSens HIV-1 QT and Amplicor HIV-1 monitor 1.5 assays. METHODS: Eighty-two clinical samples were collected and HIV viral load was determined with the above-mentioned two methods. RESULTS: The number of samples in which values obtained by NucliSens HIV-1 QT and Amplicor HIV-1 monitor 1.5 differed by <0.5 log10 RNA copies/ml and in which the viral load was undetectable accounted for 88.9 percent of the measures. The correlation coefficient between the two methods was 0.956 in 56 samples of Deltalog10 VL<0.5. CONCLUSION: The results of HIV-1 viral load determination with the two methods are highly comparable.[Abstract] [Full Text] [Related] [New Search]