These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: [Comparison of NucliSens HIV-1 QT and Amplicor HIV-1 monitor 1.5 in detecting HIV-1 viral load].
    Author: Pan PL, Xu WY, Yao J, Tao XX, Pei LJ, Jiang Y.
    Journal: Zhonghua Shi Yan He Lin Chuang Bing Du Xue Za Zhi; 2007 Jun; 21(2):177-9. PubMed ID: 17653329.
    Abstract:
    OBJECTIVE: To compare the results of detecting HIV-1 load by using NucliSens HIV-1 QT and Amplicor HIV-1 monitor 1.5 assays. METHODS: Eighty-two clinical samples were collected and HIV viral load was determined with the above-mentioned two methods. RESULTS: The number of samples in which values obtained by NucliSens HIV-1 QT and Amplicor HIV-1 monitor 1.5 differed by <0.5 log10 RNA copies/ml and in which the viral load was undetectable accounted for 88.9 percent of the measures. The correlation coefficient between the two methods was 0.956 in 56 samples of Deltalog10 VL<0.5. CONCLUSION: The results of HIV-1 viral load determination with the two methods are highly comparable.
    [Abstract] [Full Text] [Related] [New Search]