These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: An unusual cause of oversensing leading to inappropriate ICD discharges. Author: Spencker S, Poppelbaum A, Müller D. Journal: Int J Cardiol; 2008 Sep 16; 129(1):e24-6. PubMed ID: 17669521. Abstract: A 70 year old patient was treated with a cardioverter-defibrillator because of repetitive syncope and sustained ventricular tachycardia. Six months after implantation 49 shocks were delivered by the ICD while she was cleaning her apartment. Upon interrogation p-wave oversensing during sinus tachycardia was found to be responsible for the inappropriate shocks. The cause of the oversensing was a lead dislodgment caused by a Twiddler s Syndrome. As it turned out in the presented case, the main causes for the Twiddler s Syndrome are patients manipulate the devices and surgeons create oversized pockets. Additional predisposing factors are: psychiatric disease, weight loss, obesity, advanced age, female gender and a loop of the lead outside the pocket. The use of subfascial instead of submuscular ICD pockets may reduce the frequency of this complication but it is still crucial to avoid oversized pockets and to instruct the patients not to "play" with their device.[Abstract] [Full Text] [Related] [New Search]