These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Quantitation of hypoxanthine in plasma from patients with ischemic heart disease: adaption of a high-performance liquid chromatographic method. Author: Himmel HM, Sadony V, Ravens U. Journal: J Chromatogr; 1991 Jul 17; 568(1):105-15. PubMed ID: 1770088. Abstract: A high-performance liquid chromatographic method is described for the separation and quantitation of several purine compounds, including hypoxanthine. The isocratic separation of a standard mixture of nine compounds is achieved within 20 min on a reversed-phase Nucleosil 100-5C18 column, with a mobile phase of KH2PO4 (300 mM, pH 4.0)-methanol-acetonitrile-tetrahydrofuran (97.9:1:1:0.1, v/v). Uric acid, guanine, hypoxanthine, uridine, xanthine, allopurinol, inosine, guanosine and 7-methylxanthine were almost completely baseline-separated, with detection limits in the range 0.5-1.2 pmol per injection. The influence of the concentrations of buffer and tetrahydrofuran on the quality of separation are described. The within-day and the day-to-day precision were satisfactory (e.g. coefficients of variation of less than 1.5 and ca. 6.0%, respectively, for peak heights). The recovery of [3H]hypoxanthine added to samples was 86 +/- 1%. Hypoxanthine was quantified in human plasma samples obtained at various times during coronary artery bypass grafting. The hypoxanthine levels measured immediately after release of the aortic cross-clamp were significantly higher than those determined under control conditions (18.8 +/- 7.0 and 3.4 +/- 1.0 microM, respectively).[Abstract] [Full Text] [Related] [New Search]