These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Determination of total cholesterol in serum by high-performance liquid chromatography with electrochemical detection.
    Author: Hojo K, Hakamata H, Ito A, Kotani A, Furukawa C, Hosokawa YY, Kusu F.
    Journal: J Chromatogr A; 2007 Sep 28; 1166(1-2):135-41. PubMed ID: 17723233.
    Abstract:
    A simple and sensitive HPLC method that does not require derivatization for determining cholesterol has been developed. Investigation of voltammetric behavior of cholesterol showed that cholesterol could be oxidized on a glassy carbon electrode in non-aqueous solvents. This was applied to the development of a method by HPLC with electrochemical detection (HPLC-ED). The HPLC-ED was optimized using the separation of cholesterol and oxysterols including 26-hydroxycholesterol and 24S-hydroxycholesterol. The separation was carried out with a Develosil C30-UG-3 column; acetonitrile-2-propanol (9:1, v/v) containing 50mM LiClO(4) as a mobile phase; and an applied potential at 1.9V versus Ag/AgCl. The current peak height was linearly related to the amount of cholesterol injected from 0.5-100 microM (r>0.999). The detection limit (S/N=3) of cholesterol was 0.36 microM (1.8 pmol). Cholesterol at 100 microM was directly detected with a relative standard deviation (RSD) of less than 1.0% (n=8). Total cholesterol and free cholesterol in control human serum were determined by the present method with the recovery of more than 90% and the RSD (n=6) of less than 3.0%.
    [Abstract] [Full Text] [Related] [New Search]