  • Title: Instrument for x-ray magnetic circular dichroism measurements at high pressures.
    Author: Haskel D, Tseng YC, Lang JC, Sinogeikin S.
    Journal: Rev Sci Instrum; 2007 Aug; 78(8):083904. PubMed ID: 17764332.
    Abstract:
    An instrument has been developed for x-ray magnetic circular dichroism (XMCD) measurements at high pressures and low temperatures. This instrument couples a nonmagnetic copper-beryllium diamond anvil cell featuring perforated diamonds with a helium flow cryostat and an electromagnet. The applied pressure can be controlled in situ using a gas membrane and calibrated using Cu K-edge x-ray absorption fine structure measurements. The performance of this instrument was tested by measuring the XMCD spectra of the Gd(5)Si(2)Ge(2) giant magnetocaloric material.