These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Instrument for x-ray magnetic circular dichroism measurements at high pressures. Author: Haskel D, Tseng YC, Lang JC, Sinogeikin S. Journal: Rev Sci Instrum; 2007 Aug; 78(8):083904. PubMed ID: 17764332. Abstract: An instrument has been developed for x-ray magnetic circular dichroism (XMCD) measurements at high pressures and low temperatures. This instrument couples a nonmagnetic copper-beryllium diamond anvil cell featuring perforated diamonds with a helium flow cryostat and an electromagnet. The applied pressure can be controlled in situ using a gas membrane and calibrated using Cu K-edge x-ray absorption fine structure measurements. The performance of this instrument was tested by measuring the XMCD spectra of the Gd(5)Si(2)Ge(2) giant magnetocaloric material.[Abstract] [Full Text] [Related] [New Search]