These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Incidence and volume of pelvic osteolysis at early follow-up with highly cross-linked and noncross-linked polyethylene. Author: Leung SB, Egawa H, Stepniewski A, Beykirch S, Engh CA, Engh CA. Journal: J Arthroplasty; 2007 Sep; 22(6 Suppl 2):134-9. PubMed ID: 17823032. Abstract: Highly cross-linked polyethylene has shown decreased wear rates as compared to conventional polyethylene. However, the impact of this decrease on the occurrence of osteolysis remains uncertain. Forty hips implanted with noncross-linked Enduron and 36 implanted with 5-Mrad cross-linked Marathon polyethylene had a computed tomography at a minimum of 5 years after arthroplasty. Polyethylene wear, osteolysis incidence, location, and volume were compared. The incidence of osteolysis was statistically greater for patients with noncross-linked Enduron (11/40, 28%) compared to patients with the moderately cross-linked Marathon (3/36, 8%; P = .04). The average lesion volume for hips with Enduron liners (7.5 +/- 6.7 cm(3)) was significantly greater than the average lesion volume for hips implanted with Marathon liners (1.2 +/- 0.1 cm(3), P = .01). Marathon cross-linked polyethylene has shown to have a decreased incidence and volume of pelvic osteolysis. Longer follow-up is necessary to determine if Marathon cross-linked polyethylene will continue to demonstrate the encouraging improved wear and osteolysis characteristics.[Abstract] [Full Text] [Related] [New Search]