These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: High-performance liquid chromatographic determination of acetone in blood and urine in the clinical diagnostic laboratory.
    Author: Brega A, Villa P, Quadrini G, Quadri A, Lucarelli C.
    Journal: J Chromatogr; 1991 Aug 16; 553(1-2):249-54. PubMed ID: 1787157.
    Abstract:
    A method for the determination of acetone in plasma or urine by high-performance liquid chromatography (HPLC) was developed. Plasma specimens are deproteinized with acetonitrile (1:1, v/v) 2,4-dinitrophenylhydrazine (DNPH) is added to the supernatant or to filtered urine samples, similarly treated with acetonitrile (2:1, v/v) to prevent crystallization of the synthesized phenylhydrazone. An aliquot (20 microliters) of the reaction mixture was subjected to HPLC at ambient temperature using a reversed-phase Pecosphere 3 x 3 C18 column with acetonitrile-water (45:55, v/v) as eluent at a flow-rate of 1 ml/min and detection at 365 nm. Hydroxyacetone and acetoacetate phenylhydrazone derivatives do not interfere. The identification of acetone by its retention time was confirmed by comparison with a laboratory-synthesized acetone DNPH derivative. The concentration of acetone, eluted within 3 min, was determined by the peak-height method. The detection limit was 0.034 mmol/l; the relative standard deviations were less than 5% within run (n = 20) and less than 10% between run (n = 20).
    [Abstract] [Full Text] [Related] [New Search]