These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Detection of methicillin resistance in Staphylococcus aureus isolated from pediatric patients: is the cefoxitin disk diffusion test accurate enough?
    Author: Mimica MJ, Berezin EN, Carvalho RL, Mimica IM, Mimica LM, Sáfadi MA, Schneider E, Caiaffa-Filho HH.
    Journal: Braz J Infect Dis; 2007 Aug; 11(4):415-7. PubMed ID: 17873996.
    Abstract:
    We evaluated the performance of several methods for the detection of methicillin resistance in Staphylococcus aureus using 101 clinical S. aureus isolates from pediatric patients in a tertiary hospital in Brazil; 50 isolates were mecA-positive and 51 were mecA-negative. The Etest and oxacillin agar screening plates were 100% sensitive and specific for mecA presence. Oxacillin and cefoxitin disks gave sensitivities of 96 and 92%, respectively, and 98% specificity. Alterations of CLSI cefoxitin breakpoints increased sensitivity to 98%, without decreasing specificity. Our results highlight the importance of a continuing evaluation of the recommended microbiological methods by different laboratories and in different settings. If necessary, laboratories should use a second test before reporting a strain as susceptible, especially when testing strains isolated from invasive or serious infections. With the new (2007) CLSI breakpoints, the cefoxitin-disk test appears to be a good option for the detection of methicillin resistance in S. aureus.
    [Abstract] [Full Text] [Related] [New Search]