  • Title: Analytic evaluation of the beta-human chorionic gonadotropin assay on the Abbott IMx and Elecsys2010 for its use in doping control.
    Author: Shahzad K, Kim DH, Kang MJ.
    Journal: Clin Biochem; 2007 Nov; 40(16-17):1259-65. PubMed ID: 17884034.
    Abstract:
    OBJECTIVES: The principal objective of this study was to compare the analytical performance of the Elecsys2010 (Roche Diagnostics) system with the IMx (Abbott laboratories) system for beta-hCG assay in order to assess its possible utility as a confirmation test for the quantitative measurement of beta-hCG in urine for doping control purposes. DESIGN AND METHODS: Urine samples with spiked standard known concentrations of beta-hCG and different urine samples from athletes were used in order to determine the calibration curve stability and linearity, detection limit, total, within-run and between-run precision, and method comparison for the IMx and Elecsys2010 systems for beta-hCG assay, along with the stability of samples, at room temperature and at 4 degrees C. RESULTS: The IMx assay was linear up to 500 IU/L, whereas the Elecsys2010 assay was linear up to 1000 IU/L. The detection limit for the IMx and Elecsys2010 systems were 0.75 IU/L and 0.25 IU/L, respectively. The total precision of the IMx and Elecsys2010 systems were <or=5% for beta-hCG concentrations ranging from 20 to 900 IU/L and from 6.67 to 55.2 IU/L, respectively. The within-run precision tests for the IMx and Elecsys2010 systems yielded CV results of 2-3% and 3-4% respectively, while the between-run precision for the Elecsys2010 was less than 4%, whereas that of IMx was 9-16%. Regression analysis with spiked urine samples yielded the following: slope (SD) 0.7104 (0.003) and y-intercept (SD) -0.4286 (1.397) (Sy.x=2.328; r=0.9999; n=5), whereas regression analysis using the athletes' urine samples yielded the following results: slope (SD) 0.3692 (0.003) and y-intercept (SD) 0.0016 (0.003) (Sy.x=0.0096; r=0.9924; n=100). Sample stability studies evidenced a nonsignificant effect at 4 degrees C and room temperature (25 degrees C) for up to 1 month. CONCLUSIONS: For doping control, the IMx and Elecys2010 may be used to screen and confirm the presence of beta-hCG in urine of athletes, respectively. Additional studies are needed to verify this finding. This study also underscores the need for beta-hCG assays standardization for doping control.
    [Abstract] [Full Text] [Related] [New Search]