These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Long intraaortic balloon treatment time leads to more vascular complications. Author: Christenson JT, Sierra J, Romand JA, Licker M, Kalangos A. Journal: Asian Cardiovasc Thorac Ann; 2007 Oct; 15(5):408-12. PubMed ID: 17911069. Abstract: Intraaortic balloon counterpulsation is an established and efficient therapy. Limb ischemia is the most common complication. The impact of treatment duration on balloon-related complications was analyzed retrospectively in 135 patients who underwent balloon counterpulsation between 1998 and 2004. Thirty high-risk coronary patients required preoperative intraaortic balloon therapy, 41 were in preoperative cardiogenic shock, and 64 needed support for difficulties in weaning from cardiopulmonary bypass. No balloon-related mortality occurred. The overall balloon-related complication rate was 20/135 (14.8%); 18 had limb ischemia, of whom 6 (4.4%) required vascular interventions. Intraaortic balloon treatment time was significantly longer in patients who developed limb ischemia (99.8 +/- 54.1 h) compared to those who did not (34.4 +/- 30.4 h). Preoperative therapy had short treatment times and few complications. Intraaortic balloon pumping provides effective circulatory support with a low complication rate. A clear relationship was established between duration of treatment and balloon-related complications. Independent risk factors for balloon-related complications were long treatment time, acute myocardial infarction, age over 65 years, and ejection fraction less than 0.30.[Abstract] [Full Text] [Related] [New Search]