These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: A novel flow-through fluorescence optosensor for the determination of thiabendazole.
    Author: Piccirilli GN, Escandar GM.
    Journal: Anal Chim Acta; 2007 Oct 10; 601(2):196-203. PubMed ID: 17920392.
    Abstract:
    This paper presents the development of a new flow-injection system combined with solid-surface fluorescence detection for the determination of the widely used fungicide thiabendazole. Nylon powder was probed as a novel solid support for building the optosensor. The method is based on the on-line immobilization of thiabendazole onto nylon in a continuous flow system, followed by the measurement of its native fluorescence. Aqueous samples are directly injected in a water carrier, resulting in a very simple and economical method. The analytical figures of merit obtained using 1500 microL of sample and 75% methanol (v/v) as eluting solution were: linear calibration range from 8 to 120 ng mL(-1) (the lowest value corresponds to the quantitation limit), relative standard deviation, 0.9% (n=5) at a level of 64 ng mL(-1), limit of detection calculated according to 1995 IUPAC recommendations is to 2.8 ng mL(-1), and sampling rate of 14 samples h(-1). The potential interference from other agrochemicals, metal ions and common anions, and the viability of determining thiabendazole in real water samples were also evaluated.