Search MEDLINE/PubMed
Title: Calorimetric determination of kQ factors for NE 2561 and NE 2571 ionization chambers in 5 cm x 5 cm and 10 cm x 10 cm radiotherapy beams of 8 MV and 16 MV photons. Author: Krauss A, Kapsch RP. Journal: Phys Med Biol; 2007 Oct 21; 52(20):6243-59. PubMed ID: 17921583. Abstract: The relative uncertainty of the ionometric determination of the absorbed dose to water, D(w), in the reference dosimetry of high-energy photon beams is in the order of 1.5% and is dominated by the uncertainty of the calculated chamber- and energy-dependent correction factors k(Q). In the present investigation, k(Q) values were determined experimentally in 5 cm x 5 cm and 10 cm x 10 cm radiotherapy beams of 8 MV and 16 MV bremsstrahlung by means of a water calorimeter operated at 4 degrees C. Ionization chambers of the types NE 2561 and NE 2571 were calibrated directly in the water phantom of the calorimeter. The measurements were carried out at the linear accelerator of the Physikalisch-Technische Bundesanstalt. It is shown that the k(Q) factor of a single ionization chamber can be measured with a standard uncertainty of less than 0.3%. No significant variations of k(Q) were found for the different lateral sizes of the radiation fields used in this investigation.[Abstract] [Full Text] [Related] [New Search]