These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: The effect of the impedance of a thin hydrogel electrode on sensation during functional electrical stimulation. Author: Sha N, Kenney LP, Heller BW, Barker AT, Howard D, Wang W. Journal: Med Eng Phys; 2008 Jul; 30(6):739-46. PubMed ID: 17942361. Abstract: Surface functional electrical stimulation results in stimulation of cutaneous receptors and discomfort. The degree of non-uniformity of current distribution in the cutaneous layers in the vicinity of the electrode may influence the sensation experienced. This paper describes the effects on sensation of a thin, high impedance electrode designed to reduce the non-uniformity of current distribution. Sensation associated with stimulation via a self-adhesive electrode with much higher impedance than conventional electrodes was compared with a low impedance electrode in a single-blinded, crossover study. The high impedance electrode does not alter either the current at which sensation is first registered. However, at higher currents, the high impedance electrode allows 9% more current to be passed for an equivalent sensation to that experienced with the conventional electrode. A 28% decrease in discomfort with the use of the high impedance electrode was also reported.[Abstract] [Full Text] [Related] [New Search]