These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Analysis of opioid peptides by on-line SPE-CE-ESI-MS.
    Author: Hernández E, Benavente F, Sanz-Nebot V, Barbosa J.
    Journal: Electrophoresis; 2007 Nov; 28(21):3957-65. PubMed ID: 17960836.
    Abstract:
    In this study, SPE-CE-ESI-MS is explored for the preconcentration and separation of dilute solutions of six opioid peptides. First, a CE-ESI-MS methodology was developed and validated. LODs of around 1 microg/mL were obtained for all the studied peptides. For SPE-CE-ESI-MS experiments, a home-made SPE microcartridge containing a C18 sorbent was constructed near the inlet of the separation capillary. After optimizing the on-line preconcentration methodology, LODs between 10 and 0.1 ng/mL were achieved. Repeatability, reproducibility, durability of the microcartridges and linearity of the SPE-CE-ESI-MS methodology were also investigated and compared to the values obtained by CE-ESI-MS. Finally, human plasma samples fortified with opioid peptides were analyzed by SPE-CE-ESI-MS in order to show the potential of the methodology for the analysis of biological fluids.
    [Abstract] [Full Text] [Related] [New Search]