These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: [Effect of intraocular pressure measurement through therapeutic soft contact lenses by noncontact tonometer]. Author: Sugimoto-Takeuchi R, Yamamoto R, Kuwayama Y, Kinoshita S. Journal: Nippon Ganka Gakkai Zasshi; 1991 Sep; 95(9):869-72. PubMed ID: 1796769. Abstract: The measurement of intraocular pressure (IOP) was compared with and without soft contact lenses by a non-contact tonometer. The soft contact lenses examined were Plano-T and Plano-B4 therapeutic contact lenses and Breath-O refractive lens. Twenty-nine eyes of 18 patients with an IOP ranging from 9 to 48mmHg were studied. The measurements of IOP were 19.7 +/- 8.6mmHg with Plano-T and 18.9 +/- 9.2mmHg with Plano-B4. Both numbers were not statistically different, when compared with controls (19.3 +/- 9.8mmHg without lens). There was, however, a significant difference significant difference with (44.7 +/- 10.7mmHg) and without the Breath-O (p less than 0.01). The results suggest that accurate IOP measurements can be obtained through therapeutic soft contact lens by a non-contact tonometer.[Abstract] [Full Text] [Related] [New Search]